Plum A+™ Infusion System with Hospira MedNet™

Product specifications below. Click here for more product information.

Availability

Out of Stock Next Delivery: Next Delivery Date

Date product expected to arrive at Hospira's US distribution centers. Orders will release as deliveries arrive. All orders filled by Estimated Recovery date.

Devices may be shipped via Medical Necessity
Hospira Product Numbers List Number20792
Case Information Order Quantity1 Unit Case Size1 x 1 Cases per Tier12 Cases per Pallet48
Pump LinePlum™ A+
Pump TypeGeneral Infusion
DimensionsApprox. 8" h x 8" w x 6" d (excluding pole clamp protrusion and power cord storage)
WeightApprox. 9.5 lbs. (with battery)
Battery LifeApprox. 3 hours at 125 mL/hr or less, or delivers 250 mL if > 126 mL/hr
Power Requirements 120 V, 50-60 Hz, 50 VA. Meets UL 60601-1
Recharge Time Approx. 6 hours with one line operating at 125 mL/hr
Volume Range 0.1 to 99.9 mL (in 0.1 mL increments), 100 to 9999 mL (in mL increments)
Delivery Rate Accuracy ± 5% (1-999 mL/hr), ± 10% (0.1-0.9 mL/hr)
Delivery Rates Primary: 0.1 mL/hr Minimum, 999 mL/hr Maximum
Secondary: 0.1 mL/hr Minimum, 999 mL/hr Maximum
Concurrent: 0.5 mL (per line) Minimum, 500 mL/hr (cumulative) Maximum
Delivery Units mL/hr, mcg/kg/min, mcg/min, mcg/kg/hr, cg/hr, mg/min, mg/kg/hr, mg/hr, ng/kg/min, g/hr, mEq/hr, Million units/hr, units/min, units/hr, units/kg/hr, units/kg/min, mmol/min, mmol/hr
System Alarms Air-in-Line, Air-in-Line Backpriming, Low Battery, Occlusion, Turn to Run, Flow Detector, VTBI Complete or Dose End
Occlusion Alarm Pressure 1-15 psig (default 6 psig)
KVO Rate 1.0 mL/hr or the last primary delivery rate, whichever is less