Plum 360™ Infusion System with Hospira MedNet™

Product specifications below.  Click here for more product information.
Availability In Stock
Hospira Product Numbers List Number 30010
Case Information Order Quantity 1 Unit Case Size 1 x 1
Pump LinePlum 360™
Pump TypeGeneral Infusion
DimensionsApprox. 8” H x 8” W x 6” D (0.20 m H x 0.20 m W x 0.15 m), excluding pole clamp extrusion and power cord storage
WeightApprox. 10 lbs. (4.5 kg) with battery
Battery Life Approx. 3 hours at up to 125 mL/hr on a single line, or delivers at least 250 mL if > 126 mL/hr
Power Requirements 120 V~, 50 to 60 Hz, 50 VA
Recharge Time Approx. 8 hours with the device operating at 125 mL/hr on one line
Volume Range
0.1 to 99.9 mL (in 0.1 mL increments), 100 to 9999 mL (in 1 mL increments)
Delivery Rate Accuracy ± 5% (1 to 999 mL/hr), ± 10% (0.1 to 0.9 mL/hr)
Delivery Rates

Lines A and B:  0.1 to 99.9 mL/hr (in 0.1 mL increments), 100 to 999 mL/hr (in 1 mL increments)

Concurrent Delivery:  0.5 mL/hr minimum for each line

PlumSet:  500 mL/hr cumulative (A+B) maximum

KVO:  1.0 mL/hr or the last delivery rate on the associated line, whichever is less

Delivery Units mL/hr, mcg/kg/min, mcg/min, mcg/kg/hr, mg/min, mg/m2, mg/m2/min, mg/m2/hr, mg/kg, mg/kg/hr, mg/hr, ng/kg/min, g/hr, mEq/hr, mUn/min, units/min, units/hr, units/kg/hr
System Alarms Occlusion, Air-in-Line, Low Battery, Door Opened While Pumping, Lockout Violation, VTBI Complete, Valve/Cassette Test Failure, Nurse Call, No Action, Programming Not Complete
Occlusion Alarm Pressure Distal Pressure Limit (without alarm):  1-15 psig (default 6 psig)